DS-CLAMP-200DC | DS-CLAMP-500DCS | DS-CLAMP-500DC | DS-CLAMP-1000DC | |||||
Type | Flux gate | Flux gate | Flux gate | Flux gate | ||||
Current Range | 200 A DC or AC RMS | 500 A DC or AC RMS | 500 A DC or AC RMS | 1000 A DC or AC RMS | ||||
Sensitivity | ± 10 mV/A | ± 4 mV/A | ± 4 mV/A | ± 2 mV/A | ||||
Bandwidth | DC to 500 kHz | DC to 200 kHz | DC to 100 kHz | DC to 20 kHz | ||||
TEDS | Fully supported | Fully supported | Fully supported | Fully supported | ||||
Conductor Position Sensitivity |
max. ± 0.1 % of reading | max. ± 0.1 % of reading | max. ± 0.2 % of reading | max. ± 0.2 % of reading | ||||
Zero Offset (+25°C) | max. ± 10 mA | max. ± 250 mA | max. ± 250 mA | max. ± 50 mA | ||||
Error due Earth Magnetic Field |
max. ± 50 mA | max. ± 100 mA | max. ± 150 mA | max. ± 150 mA | ||||
Temp. Coefficient | ± 0.01 %/°C | ± 0.01 %/°C | ± 0.01 %/°C | ± 0.01 %/°C | ||||
Accuracy | Amplitude | Phase Error | Amplitude | Phase Error | Amplitude | Phase Error | Amplitude | Phase Error |
DC | ± 0.3 % rdg. ± 40 mA |
± 0.3 % rdg. ± 100 mA |
± 0.3 % rdg. ± 100 mA |
± 0.3 % rdg. ± 200 mA |
||||
to 100 Hz | ± 0.3 % rdg. ± 20 mA |
± 0.1 ° | ± 0.3 % rdg. ± 50 mA |
± 0.1 ° | ± 0.3 % rdg. ± 50 mA |
± 0.1 ° | ± 0.3 % rdg. ± 100 mA |
± 0.1 ° |
100 - 500 Hz | ± 0.3 % rdg. ± 40 mA |
± 0.2 ° | ± 0.3 % rdg. ± 100 mA |
± 0.2 ° | ± 0.3 % rdg. ± 100 mA |
± 0.2 ° | ± 0.5 % rdg. ± 200 mA |
± 0.2 ° |
500 Hz - 1 kHz | ± 0.5 % rdg. ± 40 mA |
± 0.5 ° | ± 0.5 % rdg. ± 100 mA |
± 0.5 ° | ± 0.5 % rdg. ± 100 mA |
± 0.5 ° | ± 1.0 % rdg. ± 200 mA |
± 0.5 ° |
1 kHz - 5 kHz | ± 1.0 % rdg. ± 40 mA |
± 1.0 ° | ± 1.0 % rdg. ± 100 mA |
± 1.0 ° | ± 1.0 % rdg. ± 100 mA |
± 1.5 ° | ± 2.0 % rdg. ± 200 mA |
± 1.5 ° |
5 kHz - 10 kHz | ± 1.5 % rdg. ± 40 mA |
± 1.5 ° | ± 1.5 % rdg. ± 100 mA |
± 1.5 ° | ± 1.5 % rdg. ± 100 mA |
± 2.0 ° | ± 5.0 % rdg. ± 500 mA |
± 2.0 ° |
10 kHz - 20 kHz | ± 5 % rdg. ± 40 mA |
± (0.5 + 0.1 f [kHz]) ° |
± 5 % rdg. ± 100 mA |
± (0.5 + 0.1 f [kHz]) ° | ± 5 % rdg. ± 100 mA |
± (0.2 f [kHz]) ° | ± 30 % rdg. ± 1000 mA |
± 10.0 ° |
20 kHz - 50 kHz | ± 5 % rdg. ± 40 mA |
± (0.5 + 0.1 f [kHz]) ° |
± 5 % rdg. ± 100 mA |
± (0.5 + 0.1 f [kHz]) ° | ± 10 % rdg. ± 250 mA |
± (0.2 f [kHz]) ° | ± 30 % rdg. ± 1000 mA |
± 10.0 ° |
50 kHz - 100 kHz | ± 15 % rdg. ± 100 mA |
± (0.5 + 0.1 f [kHz]) ° |
± 15 % rdg. ± 250 mA |
± (0.5 + 0.1 f [kHz]) ° | ± 30 % rdg. ± 250 mA |
± (0.2 f [kHz]) ° | ± 30 % rdg. ± 1000 mA |
± 10.0 ° |
100 kHz - 200 kHz | ± 5 % rdg. ± 40 mA |
± (0.5 + 0.1 f [kHz]) ° |
± 30 % rdg. ± 250 mA |
± (0.5 + 0.1 f [kHz]) ° | / | / | / | / |
200 kHz - 300 kHz | ± 5 % rdg. ± 40 mA |
± (0.5 + 0.1 f [kHz]) ° |
/ | / | / | / | / | / |
300 kHz - 500 kHz | ± 30 % rdg. ± 100 mA |
/ | / | / | / | / | / | |
Dimensions [mm] | 153 x 67 x 25 | 153 x 67 x 25 | 238 x 114 x 35 | 238 x 114 x 35 | ||||
Max. Conductor Size | 20 mm | 20 mm | 50 mm | 50 mm | ||||
Power Supply | ± 11 V to ± 15 V | ± 11 V to ± 15 V | ± 11 V to ± 15 V | ± 11 V to ± 15 V | ||||
Current Consumption | max. 6 VA | max. 7 VA | max. 7 VA | max. 7 VA | ||||
Operating Environment | Indoors, Pollution Degree II, altitude up to 2000m |
Indoors, Pollution Degree II, altitude up to 2000m |
Indoors, Pollution Degree II, altitude up to 2000m |
Indoors, Pollution Degree II, altitude up to 2000m |
||||
Operating Temp. | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C | ||||
Operating Humidity | up to 80 % (not condensing) | up to 80 % (not condensing) | up to 80 % (not condensing) | up to 80 % (not condensing) | ||||
Safety Standards | EN61010-2-032:2012 | EN61010-2-032:2012 | EN61010-2-032:2012 | EN61010-2-032:2012 | ||||
EMC Standards | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 | ||||
Rohs | EN50581:2012 | EN50581:2012 | EN50581:2012 | / | ||||
PWR-MCTS2 needed | Yes | Yes | Yes | Yes | ||||
Compatible amplifiers | SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
||||
Cable length | 3 m - MCTS to Clamp 0.3 m - Amplifier to MCTS |
3 m - MCTS to Clamp 0.3 m - Amplifier to MCTS |
3 m - MCTS to Clamp 0.3 m - Amplifier to MCTS |
3 m - MCTS to Clamp 0.3 m - Amplifier to MCTS |
||||
Data Sheet | DS-CLAMP-200DC | DS-CLAMP-500DCS | DS-CLAMP-500DC | DS-CLAMP-1000DC |
DS-CLAMP-150DC | DS-CLAMP-150DCS | DS-CLAMP-1800DC | |
Type | Hall Effect | Hall Effect | Hall Effect |
Current Range | 200 A DC or 150 A AC RMS | 250 A DC or 150 A AC RMS | 1800 A DC or AC RMS |
Sensitivity | 20 mV/A | 20 mV/A | 1 mV/A |
Accuracy (+25°C) | 1 % + 2 mA | 1 % + 2 mA | 0 - 1000 A: ± 2.5 % of reading ± 0.5 A 1000 - 1500 A: ± 3.5 % of reading 1500 - 1800 A: ± 5 % of reading |
Bandwidth | DC to 100 kHz | DC to 100 kHz | DC to 20 kHz |
Resolution | ± 1 mA | ± 1 mA | ± 1 mA |
TEDS | Fully supported | Fully supported | Fully supported |
Conductor Position Sensitivity |
± 0.5 % | ± 0.5 % | ± 1.5 % |
Zero Offset (+25°C) | ± 10 mV max. + 1 mV/°C |
± 10 mV max. + 1 mV/°C |
|
Error du Earth Magnetic Field |
± 0.5 mV | ±0.5 mV | max. 150 mA |
Temp. Coefficient | ± 0.02 % of reading / °C | ± 0.02 % of reading / °C | ± 0.15 % of reading / °C |
Zero | Auto zero @ power supply of sensor |
Auto zero @ power supply of sensor |
Auto zero via push-button |
Dimensions [mm] | 205 x 60 x 15 | 106 x 100 x 25 | 205 x 60 x 15 |
Max. Conductor Size |
32 mm | 25 mm | 32 mm |
Power Supply | ± 10 V | ± 10 V | + 9 V |
Current Consumption |
30 mA + 1mA/A measured | 30 mA + 1mA/A measured | max. 25mA |
Jaws Open Indication |
Yes | No | No |
Operating Temp. | 0 to + 60 °C | 0 to + 60 °C | 0 to + 50 °C |
Operating Humidity | 15 % to 85 % (not condensing) |
15 % to 85 % (not condensing) |
15 % to 85 % (not condensing) |
Safety Standards | EN 61010-1:2010 EN 61010-2-032:2012 EN 61010-2-031:2012 |
EN 61010-1:2010 EN 61010-2-032:2012 EN 61010-2-031:2012 |
EN 61010-1:2010 EN 61010-2-032:2012 EN 61010-2-031:2012 |
Safety | 300 Vrms CAT III | 300 Vrms CAT III | 300 Vrms CAT III |
EMC Standards | EN 61326-2-2:2013 | EN 61326-2-2:2013 | EN 61326-2-2:2013 |
PWR-MCTS2 needed |
No | No | No |
Compatible Amplifiers |
SIRIUS LV / HS-LV / XHS-LV | SIRIUS LV / HS-LV / XHS-LV | SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG SIRIUS STGM KRYPTON STG, IOLITE STG DEWE-43 |
Cable length | 3 m | 3 m | 3 m |
Data Sheet | DS-CLAMP-150DC | DS-CLAMP-150DCS | DS-CLAMP-1800DC |
DS-CLAMP-5AC | DS-CLAMP-15AC | DS-CLAMP-200AC | DS-CLAMP-1000AC | |
Type | Iron-Core | Iron-Core | Iron-Core | Iron-Core |
Range | 5 A AC RMS | 15 A AC RMS | 200 A AC RMS | 1000 A AC RMS |
Sensitivity | 100 mV/A | 100 mV/A | 10 mV/A | 1 mV/A |
Accuracy (+25 °C) | 5 mA - 0.5 A ± 2 % of reading 0.5 A - 1 A ± 1 % of reading 1 A - 12 A ±0 .5 % of reading |
< 1 A ± 2.5 % of reading 1 A - 15 A ± 1 % of reading |
0.5 A - 10 A ± 3.5 % of reading 10 A - 100 A ± 2.5 % of reading 100 A - 240 A ± 1 % of reading |
< 1 A ± 2 % of reading 10 A - 100 A ± 0.5 % of reading 100 A - 1200 A ± 0.3 % of reading |
Phase Error | 5 mA - 0.5 A ± 2 ° 0.5 A - 1 A ± 1° 1 A - 12 A ± 1 ° |
< 1 A ± 5 ° 1 A - 15 A ± 3 ° |
0.5 A - 10 A not specified 10 A - 100 A ≤ 5 ° 100 A - 240 A ≤ 2.5 ° |
< 1 A not specified 10 A - 100 A ≤ 1 ° 100 A - 1200 A ≤ 0.7 ° |
Bandwidth | 2 Hz to 5 kHz | 2 Hz to 10 kHz | 2 Hz to 10 kHz | 2 Hz to 10 kHz |
Resolution | 0.01 A | 0.01 A | 0.5 A | 0.001 A |
TEDS | Fully supported | Fully supported | Fully supported | Fully supported |
Overload Capability | 1.5 x I nominal | 1.3 x I nominal | 1.3 x I nominal | 1.3 x I nominal |
Conductor Position Sensitivity |
≤ 0.5 % of reading | ≤ 0.5 % of reading | ≤ 0.5 % of reading | ≤ 0.1 % of reading |
Influence of Crest Factor |
≤ 1 % | ≤ 3 % | ≤ 3 % | ≤ 1 % |
Error due Earth Magnetic Field |
/ | ≤ 15 mV/A @ 50 Hz | ≤ 15 mV/A @ 50 Hz | ≤ 15 µV/A @ 50 Hz |
Temperature Coefficient |
±0.015 % / °C | ≤ 0.2 % / 10 °C | ≤ 0.2 % / 10 °C | ≤ 0.1 % / 10 °C |
Influence of Frequency |
/ | < 5 % of output signal from 20 Hz ... 1 kHz < 10 % of output signal from 1 kHz ... 10 kHz |
< 3 % of output signal from 20 Hz ... 1 kHz < 12 % of output signal from 1 kHz ... 10 kHz |
< 1 % of output signal from 30 Hz ... 48 Hz < 0.5 % of output signal from 35 Hz ... 1 kHz < 1 % of output signal from 1 kHz ... 5 kHz |
Dimensions [mm] | 102 x 34 x 24 | 135 x 51 x 30 | 135 x 51 x 30 | 216 x 111 x 45 |
Conductor Diameter |
15 mm | 20 mm | 20 mm | 52 mm |
Operating Temperature |
-10 °C to +55 °C | -40 °C to +70 °C | -40 °C to +70 °C | -40 °C to +70 °C |
Operating Humidity |
up to 85 % (not condensing) |
up to 85 % (not condensing) | up to 85 % (not condensing) | up to 85 % (not condensing) |
Safety Standards | EN61010-2 EN61010-2-032:2012 |
EN61010-2 EN61010-2-032:2012 |
EN61010-2 EN61010-2-032:2012 |
EN61010-2 EN61010-2-032:2012 |
Safety | CAT III 600V CAT IV 300V |
CAT III 600V CAT IV 300V |
CAT III 600V CAT IV 300V |
CAT III 600V CAT IV 300V |
EMC Standards | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 |
PWR-MCTS2 needed | No | No | No | No |
Compatible Amplifiers |
SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
Cable length | 2 m | 1.4 m | 1.4 m | 1.4 m |
Data Sheet | DS-CLAMP-5AC | DS-CLAMP-15AC | DS-CLAMP-200AC | DS-CLAMP-1000AC |
CT 100 | CT 200 | CT 400 | CT 500 | CT 1000 | CT 2000 | |
Type | Zero-Flux | Zero-Flux | Zero-Flux | Zero-Flux | Zero-Flux | Zero-Flux |
Primary Current Range DC RMS Sinus |
100 A | 200 A | 400 A | 500 A | 1000 A | 2000 A |
Conversion ratio | 1:500 | 1:1000 | 1:1500 | 1:750 | 1:1500 | 1:2000 |
Overload Ability Short Time (100 ms) |
20 kApk | 20 kApk | 20 kApk | 5000 Apk | 5000 Apk | 10000 Apk |
Max. burden resistor (100 % of Ip) | 25 Ω | 20 Ω | 10 Ω | 2.5 Ω | 4 Ω | 3.5 Ω |
di/dt (accurately followed) | 100 A/μs | 100 A/μs | 100 A/μs | 100 A/μs | 100 A/μs | 100 A/μs |
Temperature influence | 0.2 ppm/K | 0.2 ppm/K | 0.1 ppm/K | < 0.2 ppm/K | < 0.3 ppm/K | < 0.1 ppm/k |
Nominal secondary current | 200 mA at 100 A | 200 mA at 200 A | 266 mA at 400 A | 666 mA at 500 A | 666 mA at 1000 A | 1 A at 2000 A |
Bandwidth ( -3 dB) | DC ... 2 MHz | DC ... 1.1 MHz | DC ... 800 kHz | DC ... 520 kHz | DC ... 440 kHz | DC ... 140 kHz |
Linearity | < 0.002 % | < 0.002 % | < 0.005 % | < 0.003 % | < 0.003 % | < 0.003 % |
Offset | < 0.0015 % | < 0.0010 % | < 0.0008 % | < 0.001 % | < 0.0012 % | < 0.0012 % |
Rated isolation voltage RMS, single isolation CAT III, pollution deg. 2, IEC 61010-1 standards, EN 50178 standards |
1000 V |
1000 V |
1000 V |
1000 V | 1000 V | 1000 V |
Test voltage 50 / 60 Hz, 1 min | 4.95 kV | 4.95 kV | 4.95 kV | 4.2 kV | 4.2 kV | 6 kV |
Inner diameter | 28 mm | 28 mm | 28 mm | 38 mm | 38 mm | 70 mm |
Dimensions | 93 x 78.8 x 49.5 | 94 x 78.8 x 49.5 | 95 x 78.8 x 49.5 | 128 x 106 x 54 | 128 x 106 x 54 | 231 x 220 x 76 |
Supply voltage (±5 %) | ±15 V | ±15 V | ±15 V | ±15 V | ±15 V | ±15 V |
Operating humidity | 20 % to 80 % (not condensing) |
20 % to 80 % (not condensing) |
20 % to 80 % (not condensing) |
20 % to 80 % (not condensing) |
20 % to 80 % (not condensing) |
20 % to 80 % (not condensing) |
Operating temperature | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C | - 40 °C to + 85 °C |
DEWESoft® Shunt | 5 Ω | 5 Ω | 5 Ω | 1 Ω | 1 Ω | 1 Ω |
PWR-MCTS2 needed | Yes | Yes | Yes | Yes | Yes | Yes |
Compatible amplifiers | SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
Cable length | 5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
5 m - MCTS to Transducer 0.3 m - Amplifier to MCTS |
Data Sheet |
DS-FLEX-3000-17 | DS-FLEX-3000-35 | DS-FLEX-3000-35HS | DS-FLEX-3000-80 | DS-FLEX-30000-120 | |
Type | Rogowski coil | Rogowski coil | Rogowski coil | Rogowski coil | Rogowski coil |
Current Range | 3 A, 30 A, 300 A, 3000 A AC RMS |
3 A, 30 A, 300 A, 3000 A AC RMS |
3000 A AC RMS |
3 A, 30 A, 300 A, 3000 A AC RMS |
30 A, 300 A, 3000 A, 30000 A AC RMS |
Sensitivity | 1000 mV/A, 100 mV/A, 10 mV/A, 1 mV/A |
1000 mV/A, 100 mV/A, 10 mV/A, 1 mV/A |
10 mV/A | 1000 mV/A, 100 mV/A, 10 mV/A, 1 mV/A |
100 mV/A, 10 mV/A, 1 mV/A, 0.1 mV/A |
Accuracy (+25 °C) |
≤ 1.5 % | ≤ 1.5 % | ≤ 1.0 % | ≤ 1.5 % | ≤ 1.5 % |
Phase Error | ≤ 1.0 deg | ≤ 1.0 deg | ≤ 1.0 deg | ≤ 1.0 deg | ≤ 1.0 deg |
Bandwidth | 3 A: 10 Hz to 10 kHz
Others: 10 Hz to 20 kHz |
3 A: 10 Hz to 10 kHz
Others: 10 Hz to 20 kHz |
5 Hz - 1MHz | 3 A: 10 Hz to 10 kHz
Others: 10 Hz to 20 kHz |
30 A: 10 Hz to 5 kHz Others: 10 Hz to 20 kHz |
TEDS | Not supported | Not supported | Fully supported | Not supported | Not supported |
Noise | ≤ 0.04 Arms | ≤ 0.04 Arms | ≤ 0.04 Arms | ≤ 0.04 Arms | ≤ 0.04 Arms |
Temperature Coefficient |
≤ 0.5 % / 10 °C | ≤ 0.5 % / 10 °C | ≤ 0.5 % / 10 °C | ≤ 0.5 % / 10 °C | ≤ 0.5 % / 10 °C |
Casing Dimensions [mm] |
120 x 58 x 36 | 120 x 58 x 36 | 140 x 64 x 28 | 120 x 58 x 36 | 120 x 58 x 36 |
Coil Length | 170 mm | 350 mm | 350 mm | 800 mm | 1200 mm |
Coil Diameter | 45 mm | 100 mm | 100 mm | 250 mm | 380 mm |
Power Supply | +5 V | +5 V | +9 V | +5 V | +5 V |
Operating Temperature |
- 10 °C to + 55 °C | - 10 °C to + 55 °C | - 10 °C to + 55 °C | - 10 °C to + 55 °C | - 10 °C to + 55 °C |
Operating Humidity |
up to 90 % (not condensing) |
up to 90 % (not condensing) |
up to 90 % (not condensing) |
up to 90 % (not condensing) |
up to 90 % (not condensing) |
Safety Standards |
EN60529 EN61010-2 EN61010-2-032:2012 |
EN60529 EN61010-2 EN61010-2-032:2012 |
EN60529 EN61010-2 EN61010-2-032:2012 |
EN60529 EN61010-2 EN61010-2-032:2012 |
EN60529 EN61010-2 EN61010-2-032:2012 |
Safety | CAT IV 600V | CAT IV 600V | CAT III 1000V | CAT IV 1000V | CAT IV 1000V |
EMC Standards | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 | EN61326-1:2013 |
PWR-MCTS2 needed | No | No | No | No | No |
Compatible amplifiers | SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV SIRIUS STG / HS-STG |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
SIRIUS LV / HS-LV / XHS-LV
SIRIUS STG / HS-STG SIRIUS STGM DEWE 43 |
Cable length | 2.5 m (Extendable to 10 m upon request) |
2.5 m (Extendable to 10 m upon request) |
2.5 m (Extendable to 10 m upon request) |
2.5 m (Extendable to 10 m upon request) |
2.5 m (Extendable to 10 m upon request) |
Data Sheet | DS-FLEX-3000-17 | DS-FLEX-3000-35 | DS-FLEX-3000-35HS | DS-FLEX-3000-80 | DS-FLEX-30000-120 |
Power supply unit technical specifications
SIRIUSi-PWR-MCTS2 / SIRIUSir-PWR-MCTS | SIRIUSi-HD-PWR-MCTS2 / SIRIUSir-HD-PWR-MCTS |
|
Power supply | 9 - 36 V DC | 9 - 36 V DC |
Max power consumption | 85 W | 85 W |
Physical dimensions | 265 x 140 x 65 mm | 265 x 140 x 65 mm |
Operating temperature | -20 °C to 50 °C | -20 °C to 50 °C |
Storage temperature | -40 °C to 85 °C | -40 °C to 85 °C |
Humidity (@ 60 °C) | 95 % RH non-condensing | 95 % RH non-condensing |
Output | 4 x Isolated Power supply (1500 V DC, 60 sec) | 8 x Isolated Power supply (1500 V DC, 60 sec) |
Output voltage | ±15 V DC | ±15 V DC |
Maximum output per channel | 20 W | 20 W |
Short circuit protection | indefinite (automatic recovery) | indefinite (automatic recovery) |
Overload protection | 150 % of Iout max. typ | 150 % of Iout max. typ |